WARNING
WARNING
WARNING
CAUTION
CAUTION
CAUTION
Safety and Regulatory Information
Safety Information
These servicing instructions are for use by qualified personnel only. To avoid electrical shock, do not perform any servicing unless you are qualified to do so.
The opening of covers or removal of parts is likely to expose dangerous voltages. Disconnect the instrument from all voltage sources while it is being opened.
This product is designed for use in Installation Category II and Pollution Degree 2 per IEC 1010 and 664 respectively.
Cautions
Cautions applicable to this instrument are:
Always use the
This instrument has autoranging line voltage input; be sure the supply voltage is within the specified range.
Ventilation Requirements: When installing the instrument in a cabinet, the convection into and out of the instrument must not be restricted. The ambient temperature (outside the cabinet) must be less than the maximum operating temperature of the instrument by 4 °C for every 100 watts dissipated in the cabinet. If the total power dissipated in the cabinet is greater than 800 watts, forced convection must be used.
Agilent Technologies Z5623A Option H48 |