Calibrating the Network Analyzer

Performing the Calibration

Figure 4-4 Cal Type Window When It Is First Opened

Step 6. Click the Full SOLT 2-Portbutton. The result is shown in Figure 4-5.

Figure 4-5 Cal Type Window After Selecting “Full SOLT 2-Port”

Step 7. Click the OK button. The Cal Type window will then close and the Cal Wizard window will pop back up as shown in Figure 4-6.

Figure 4-6 Cal Wizard Menu After Selection Of Cal Type Is Complete

Step 8. Verify that the currently selected calibration type shown in the Cal Wizard window is “Full SOLT 2-Port (Omit Isolation)” as shown in Figure 4-6.

Step 9. Click the Measure ECal Stds button in the Cal Wizard window. The window shown in Figure 4-7will then pop up.