Agilent Technologies Z5623A manual Troubleshooting the Front Panel Board

Models: Z5623A

1 138
Download 138 pages 46.21 Kb
Page 128
Image 128

Service

Troubleshooting

Troubleshooting the Front Panel Board

Turn the instrument power on and check the following:

Check the condition of each of the switching paths by issuing commands to switch each of the paths to either the transmission or reflection path. Ensure that the LCD indicates the appropriate path.

If the LCD indicates a wrong path, the problem can lie with either the front panel board or the main switch board. Measure the RF path to determine where the problem is.

Ensure that the front panel washers between the board and front panel assembly are present. Missing washers can cause erratic LCD behavior.

If the LCD does not display the proper path, check to see if the RF path has indeed been switched. If the problem lies with the front panel board, replace it.

Troubleshooting the Controller and Switch Driver Boards

Turn the instrument power on. Check the condition of each of the switching paths by issuing commands to switch each of the paths to either the Reflection or Transmission path. Check each of the RF paths for connection. If an RF path is not connected to the necessary port or terminated in 50 Ω, replace the controller and switch driver board.

8 -6

Agilent Technologies Z5623A Option H48

Page 128
Image 128
Agilent Technologies Z5623A Troubleshooting the Front Panel Board, Troubleshooting the Controller and Switch Driver Boards