3-8 Agilent Technologies Z5623AOption H48
Using the Network Analyzer to Control the Test Set
Common Errors in Manual Commands
NOTE

It is possible to issue several commands in a single line. For example,

the following string connects the Refle ction port to Test Port #2,

connects the Transmission port to Test Port #5, and sets the Control

Lines to OLOOL: “refl_02;tran_05;00001001;”<Enter>.

Common Errors in Manual Commands

This section explains several common errors.

Table 3-1 Common Errors in Manual Commands
Command(s) Issued Explanation of Error Solution
“refl_01” Missing semicolon. Send a semicolon:
“;”
“refl_01”
“tran_08;”
Missing semicolon in first
command causes both
commands to be interpreted
as a single invalid
command: “refl_01tran_08;”
Resend both
commands with a
semicolon in each:
“refl_01;”
“tran_08;”
“10101;” Three leading zeroes are
missing. (This is a binary
Control Line command. It
needs three leading zeroes.)
(Decimal Control Line
commands don’t need
leading zeroes.)
Resend the command
with leading zeroes
included:
“00010101;”