NOTE

Using the Network Analyzer to Control the Test Set

Common Errors in Manual Commands

It is possible to issue several commands in a single line. For example, the following string connects the Reflection port to Test Port #2, connects the Transmission port to Test Port #5, and sets the Control Lines to OLOOL: “refl_02;tran_05;00001001;”<Enter>.

Common Errors in Manual Commands

This section explains several common errors.

Table 3-1

Common Errors in Manual Commands

 

 

 

 

 

Command(s) Issued

Explanation of Error

 

Solution

 

 

 

 

 

“refl_01”

 

Missing semicolon.

 

Send a semicolon:

 

 

 

 

“;”

 

 

 

 

 

“refl_01”

 

Missing semicolon in first

 

Resend both

 

 

command causes both

 

commands with a

“tran_08;”

 

commands to be interpreted

 

semicolon in each:

 

 

as a single invalid

 

“refl_01;”

 

 

command: “refl_01tran_08;”

 

 

 

 

 

 

 

 

 

“tran_08;”

 

 

 

 

 

“10101;”

 

Three leading zeroes are

 

Resend the command

 

 

missing. (This is a binary

 

with leading zeroes

 

 

Control Line command. It

 

included:

 

 

needs three leading zeroes.)

 

“00010101;”

 

 

(Decimal Control Line

 

 

 

 

 

 

 

commands don’t need

 

 

 

 

leading zeroes.)

 

 

 

 

 

 

 