Agilent Technologies Z5623A manual Electronic Calibration Window

Models: Z5623A

1 138
Download 138 pages 46.21 Kb
Page 55
Image 55

Calibrating the Network Analyzer

Performing the Calibration

Figure 4-7 Electronic Calibration Window

Step 10. Click the Module button. The window shown in Figure 4-8will then pop up.

Figure 4-8 Module Connection Message

Step 11. Click the OK button. (The connections requested in Figure 4-8were made in Step 1 of “Performing the Calibration” on page 4-2.) This window will close, the window shown in Figure 4-7will pop back up, and the calibration will begin.

The network analyzer will complete the calibration with no further operator actions required. The calibration will take about 5 seconds.

Step 12. Verify that the calibration has been completed. This is shown by a check mark over the Module button as shown in Figure 4-9.

Figure 4-9 Electronic Calibration Window After Cal Is Complete

Step 13. Click the OK button shown in Figure 4-9. The window will close; the Cal Wizard window will pop back up with the following message as shown in Figure 4-10: “Done. Save your settings now by pressing OK”.

Agilent Technologies Z5623A Option H48

4-5

Page 55
Image 55
Agilent Technologies Z5623A manual Electronic Calibration Window