Performance Verification

Verifying Return Loss and Insertion Loss Specifications

Verifying Return Loss and Insertion Loss

Specifications

Measurements Using The Reflection Port

Step 1. Connect an RF cable from Port 1 of the network analyzer to Test Port 1 of the test set. Connect another RF cable from Port 2 of the network analyzer to the Reflection Port of the test set. See Figure 5-39.

Figure 5-39 Setup for Measuring Insertion Loss and Return Loss of Test Set

 

 

 

 

 

 

 

E8358A

 

 

 

 

 

 

 

(300 kHz - 9.0

GHz)

GPIB

 

 

 

 

 

 

 

 

 

 

 

Port 1

 

 

 

 

Port 2

 

 

Reflection

 

 

 

 

 

Transmission

GPIB

1

2

3

4

5

6

7

8

 

Step 2. Command the test set to connect the test port to the Reflection Port. Consult Table 3-2 on page 3-9for a list of commands.

Step 3. Recall the limit test file for insertion loss. It is listed in Table 5-3 on page 5-7.

Step 4. Read the results of the insertion loss test on the screen of the network analyzer. Record the results in Table 5-8.

Step 5. Recall the limit test file for return loss (port active). It is listed in Table 5-3 on page 5-7.

Step 6. Read the results of the return loss (port active) test on the screen of the network analyzer. Record the results in Table 5-8.

5 -36

Agilent Technologies Z5623A Option H48

Page 104
Image 104