Calibrating the Network Analyzer
Performing the Calibration
Or repeat your cal.”
Figure 4-10 Cal Wizard Window After Cal Is Complete
Step 14. Get a pen or pencil and a piece of paper. Step 16 will ask you to write down a filename.
Step 15. Click the OK button. The calibration will now be saved to disk in the following directory: C:\Program Files\Agilent\Network Analyzer\Documents. The file will have a “.cst” suffix.
Step 16. Copy down the name of the file into which the calibration is saved. The name will appear on the screen for about 10 seconds. Copy the filename, not the path. (The path is listed in Step 15 above.) A typical filename would be, “at006.cst”. Record also the date and time that the calibration was performed. (The date and time can be used to locate the file in case the filename you write down is incorrect or incomplete.)
Agilent Technologies Z5623A Option H48 |