Performance Verification

Setting Up Limit Testing

Step 2. When the Save As window comes up as shown in Figure 5-15, type a file name that includes the words, “insertion loss,” and the date of the calibration listed in Table 4-1on page 4-13. For example, if the calibration file is named, “cal_06_17_01”, an appropriate file name for this step would be, “insertion_loss_with_06_17_01_cal”. See Figure

5-16. After entering the file name, record it in Table 5-3 on page 5-7. Next, click the Save button.

Figure 5-15 “Save As...” Window

Figure 5-16 Specifying The Name Of The Limit Test File For Insertion Loss

5 -16

Agilent Technologies Z5623A Option H48

Page 84
Image 84
Agilent Technologies Z5623A manual Save As... Window