Performance Verification

Setting Up Limit Testing

Figure 5-35 The Status Bar Displayed at Bottom of Screen

Specifying Measurement Type, Title, and Scaling

Step 1. Click Trace on the menu bar. When the Trace menu appears, scroll down to the Measure button and pause momentarily. When the Measure menu appears, scroll down to the S21 button and click it. See Figure 5-1.

Step 2. Click View on the menu bar. When the View menu appears, scroll down to the Meas Display button and pause momentarily. When the Meas Display menu appears, scroll down to the Title... button and click it. See Figure 5-2.

Step 3. The Title Entry window is shown in Figure 5-3. When this window pops up, click in the Entry field.

Step 4. Type “Crosstalk” in the Entry field, then press the Enter key on the USB keyboard (or the OK button in the Title Entry window) to close the window and display the new title.

Step 5. Click Scale on the menu bar. When the Scale menu appears, scroll down to the Scale... button and click it. See Figure 5-5.

5 -32

Agilent Technologies Z5623A Option H48