Carrier 69NT40-511-199 manual To replace the recording thermometer element, Battery, Calibration

Models: 69NT40-521 69NT40-511-1 69NT40-511-199

1 138
Download 138 pages 33.05 Kb
Page 114
Image 114

raises stylus indicated temperature reading; shortening shaft (clockwise) lowers stylus reading. Then retighten set screw.

(b)Reset control at 0_C (32_F), start the refrigeration unit and repeat accuracy check. After temperature stabilization, recording thermometer should be within 0.3_C (1/2_F) limits.

6

2

5

1

3

4

1.Wind-up Key

2.Stylus

3.Set Screw

4.Pinion Shaft

5.Stylus Lifter

6.Bulb

Figure 6-19. Partlow Recording Thermometer

e.Replacing Recording Thermometer Element (Bulb and Capillary)

The element is mercury-filled and the temperature-pressure of the element controls the stylus, which moves across the chart in response to temperature changes as sensed by the bulb located in the evaporator supply air.

The element flange contains three O-rings. Care should be taken to install the new element flange without damaging the O-rings. It is possible for a mercury leak to develop at the flange if O-ring damage occurs.

The stylus will continue to fall (container temperature will actually be higher) if a leak develops in the flange, capillary or bulb.

To replace the recording thermometer element:

1.Turn unit OFF and disconnect power source.

2.Remove upper back panel. Remove bulb clamps securing bulb to unit.

3.Remove two flange screws from recording thermometer and feed capillary and element through the unit.

4.Push replacement bulb end and capillary through the unit.

5.Fill slots with silastic (RTV432, Dow Corning).

6.Attach bulb clamps tightly to bulb.

7.Connect element flange to recorder making sure hub of flange faces out to fit into the hole in instrument case (recording thermometer).

8.Rezero the recorder. (Refer to sections 6.19.a. through 6.19.d.)

9.Install inlet air grille and upper panel. Start unit and check recorder calibration.

CAUTION

Capillary tubing may be bent, but never sharper than 1/2” radius; extra care should be taken when bending adjacent to welds. The sensing bulb should never be bent, as this will affect calibration.

6.20 SAGINOMIYA RECORDING THERMOMETER

NOTE

Do not overtighten chart nut after replacing chart.

a. Battery

1.Open door and remove chart nut and platen.

2.Push voltage indicator test switch (item 2, Figure 6-20). Replace battery if voltage indicator points to the red or white zone.

b. Calibration

1.Install new chart on platen.

2.Place recorder bulb in ice bath (0 ¦ 0.2_C = 32 ¦ 0.35_F). (Remove rear upper panel to remove bulb.) Leave bulb immersed in ice bath for 10 minutes.

3.After 10 minutes, rotate the chart by hand and check the stylus indicated temperature. Do not touch stylus during the checkout procedure.

T-268-07

6-20

Page 114
Image 114
Carrier 69NT40-511-199, 69NT40-521 To replace the recording thermometer element, Saginomiya Recording Thermometer, Battery