(1)Requires Secondary Fieldbus (Pos 5), code A-F, W, 0, 6-9.

Reference Manual

Rosemount 2410

300530EN, Rev BA September 2012

Code (Pos 8)

Integral Display

 

Note

 

1

LCD

 

0

None

 

 

 

 

 

Code (Pos 9)

Power Supply

Note

P

Extended input range: 48-240 VAC at 50/60 Hz, and 24-48 VDC

 

Code (Pos 10)

Software

Note

S

Standard

 

 

 

 

 

Code (Pos 11)

Hazardous Location Certification

Note

E1

ATEX Flameproof with intrinsically safe output

 

 

 

 

 

E5

FM-US Explosionproof with intrinsically safe output

 

E6

FM-Canada Explosionproof with intrinsically safe output

 

E7

IECEx Flameproof with intrinsically safe output

 

K1

ATEX Flameproof + FM-US Explosionproof

 

 

 

 

 

K3

ATEX Flameproof + IECEx Flameproof

 

K4

FM-US Explosionproof + FM-Canada Explosionproof

 

 

 

 

 

NA

No hazardous location certification

 

Code (Pos 12)

Custody Transfer Type Approval

Note

R

OIML R85 E(7) 2008 performance certification

 

0

None

 

 

 

 

 

Code (Pos 13)

Housing

Note

A

Standard enclosure

Polyurethane-covered aluminium. IP 66/67

 

 

 

 

Code (Pos 14)

Cable/Conduit Connections

Note

1

½–14 NPT and ¾-14 NPT

Female thread. Includes 3 plugs

2

M20 x1.5 and M25x1.5 adapters

Female thread. Includes 3 plugs, and 3 adapters

G

Metal cable glands (½-14 and ¾-14 NPT)

Min. temperature -20 ˚C (-4 ˚F). ATEX / IECEx

 

 

 

Exe approved. Includes 3 plugs, and 3 glands

E

Eurofast male, ½–14 NPT and ¾-14 NPT

Includes 3 plugs

M

Minifast male, ½–14 NPT and ¾-14 NPT

Includes 3 plugs

Code (Pos 15)

Mechanical Installation

Note

P

Mounting kit for both wall and pipe installation

1-2 in. vertical or horizontal pipes

 

 

 

 

W

Mounting kit for wall installation

 

0

None

 

 

 

 

 

Code

Options – none or multiple selections are possible

Note

QT

IEC 61508 certificate and FMEDA-data(8)

 

ST

Engraved SST tag plate

 

 

 

 

 

Model Code Example: 2410 - S F R 0 3 2 1 P S E1 R A 1 P - ST

(1)Requires Secondary Fieldbus (Pos 5), code A-F, W, 0, 6-9.

(2)Requires Primary Fieldbus (Pos 4) code R or 4.

(3)Requires a separate Smart Wireless THUM™ Adapter (not included). Maximum Tankbus (FF) current is reduced to 200 mA.

(4)Requires Relay Output (Pos 6) code 0 or F.

(5)Requires Secondary Fieldbus (Pos 5), code 0 or Primary Fieldbus (Pos 4) code 4 and Secondary Fieldbus (Pos 5) code W, C, D, 8 or 9.

(6)Requires Secondary Fieldbus (Pos 5), code 0 or F.

(7)Requires a Rosemount 5900S gauge with the corresponding custody transfer approval. A Rosemount 2230 display or TankMaster is required for an approved read-out.

(8)Requires Relay Output (Pos 6) code 2 or 3 (SIL 3 or SIL 2).

A-6

Appendix A. Reference Data

Page 108
Image 108
Emerson Process Management Rosemount 2410 manual Reference Manual, Requires Secondary Fieldbus Pos 5, code A-F, W, 0