Image 88
6.2.11Testing the Relays

Reference Manual

Rosemount 2410

300530EN, Rev BA September 2012

6.2.11Testing the Relays

Table 6-2. Manual control of relay function

The Manual Control of Relay function lets you manually open or close the Rosemount 2410 Tank Hub built-in relays in order to verify the relay function. After the specified Safety Reset Time the relay automatically returns to normal mode.

To change relay state by using the TankMaster WinSetup program do the following:

1.In the TankMaster WinSetup workspace select the 2410 icon.

2.Click the right mouse button and choose Manual Control Relay.

3.Select the virtual relay functions to be tested; Virtual Relay 1, Virtual Relay 2, etc.

Up to ten virtual relay functions can be configured for a 2410 Tank Hub. See “Advanced Configuration in Winsetup” on page C-3and “Relay Output” on page C-6for more information.

4.Specify a Safety Reset Time. This value specifies the time period for the relay to stay in the test state. When the specified period of time has elapsed, the relay automatically returns to the original state. The relay will reset even if communication with the TankMaster PC would fail.

5.Choose the desired Manual Relay State. Available options are Alarm, Normal, and Toggle.

6.Click the Set button. Now the selected relay changes state for the specified number of seconds and then returns to the previous state.

6-14

Section 6. Service and Troubleshooting

Page 88
Image 88