• See the Rosemount 5900S Reference Manual

Reference Manual

300530EN, Rev BA September 2012

Rosemount 2410

Symptom

Possible cause

 

Action

 

 

 

 

Incorrect level reading from radar

Configuration error

Check configuration of the level gauge; in TankMaster

level gauge

 

 

Winsetup open Properties for the level gauge

 

 

 

associated with the tank.

 

 

• See the Rosemount 5900S Reference Manual

 

 

 

(Document no. 300520en) and the Raptor System

 

 

 

Configuration Manual (Document no. 300510EN) for

 

 

 

more information on how to configure a Rosemount

 

 

 

5900S Radar Level Gauge.

 

 

 

 

Measurement units are not compatible

If default database is loaded to the 2410 Tank Hub, do one

 

with host system

of the following:

 

 

in TankMaster WinSetup verify system units and

 

 

 

re-install the tank associated with the 2410 Tank Hub

 

 

update Holding Registers with correct measurement

 

 

 

units

 

 

 

 

 

Hardware failure

Check diagnostics information, see “Diagnostics” on

 

 

 

page 6-7.

 

 

• Contact Emerson Process Management/Rosemount

 

 

 

TankGauging service department.

 

 

 

 

No output on the 2410 integral

Hardware failure

Check the model code to verify that the 2410 was

display

 

 

ordered with the LCD display option.

 

 

Check display connection.

 

 

Check diagnostics information, see “Diagnostics” on

 

 

 

page 6-7.

 

 

• Contact Emerson Process Management/Rosemount

 

 

 

TankGauging service department.

 

 

 

 

The Error LED (red) is blinking

Various reasons such as hardware or

See “Error Messages” on page 5-5and “Error

 

software failure, communication, or

 

Messages” on page 6-26.

 

configuration error

Check Device Status input register (see “Device Status”

 

 

 

on page 6-23).

 

 

 

 

The Status LED (yellow) is

Normal operation. The yellow Status LED

See “LED” on page 5-6for more information.

blinking

blinks at a constant rate of one flash every

 

 

 

other second.

 

 

 

 

 

 

Configuration can not be saved

Write protection switch is set to the ON

Check write protection switch on the display, see “Write

 

position

 

Protection Switch” on page 6-11.

 

 

 

 

 

The 2410 is write protected in TankMaster

Check write protection in TankMaster WinSetup, see

 

WinSetup

 

“Write Protection” on page 6-10.

 

 

 

 

 

Application software installed which is

Reset holding registers to the default database setting,

 

incompatible with current holding register

 

see “Loading the Default Database” on page 6-16and

 

setup

 

restart the 2410 Tank Hub.

 

 

 

 

 

Corrupt holding registers

Reset holding registers to the default database setting,

 

 

 

see “Loading the Default Database” on page 6-16and

 

 

 

restart the 2410 Tank Hub.

 

 

 

 

Section 6. Service and Troubleshooting

6-21

Page 95
Image 95
Emerson Process Management Rosemount 2410 manual See the Rosemount 5900S Reference Manual