Reference Manual

Rosemount 2410

300530EN, Rev BA September 2012

Figure 3-10. Example of a Raptor system with several tanks daisy-chained to a Rosemount 2410

Figure 3-10 illustrates an example with a number of tanks daisy-chained to a Rosemount 2410 Tank Hub (requires multiple tanks option).

If a Raptor field device is connected to the end of the Tankbus (fieldbus segment), the built-in terminator can be used. A separate bus terminator may be used instead.

Tankbus length up to 1000 meter depending on number of devices and cable type

644Temperature Transmitter

5400 Level

Transmitter

2410 Tank Hub with intrinsically safe power supply, integrated power conditioner, and built-in terminator

2230 Display

2230 Display with built-in terminator

Tankbus

Note that the total length of the Tankbus (fieldbus segment) must be within the

FISCO specifications, see “Cable Selection for the Tankbus” on page 3-6.

3-18

Section 3. Installation