Emerson Process Management Rosemount 2410 5.5 LED, LED Start-Up Information, Reference Manual

Models: Rosemount 2410

1 144
Download 144 pages 24.51 Kb
Page 72
Image 72
5.5 LED

Reference Manual

Rosemount 2410

300530EN, Rev BA September 2012

5.5 LED

There are three Light Emitting Diodes (LED) on the Rosemount 2410 front for

 

status and error information.

Figure 5-3. The Rosemount

 

2410 has three LEDs

 

Error LED (Red) Error LED (Red) occurs, the Error LED flashes a sequence that corresponds to5.5.1LED Start-Up InformationManual background

Status LED (Yellow) Manual backgroundManual backgroundManual backgroundManual background

Power On LED (Green) Manual backgroundManual background

The following color codes are used for the 2410 LEDs:

Table 5-4. LED color codes

LED Type

Color

Description

 

 

 

Power On

Green

The green LED indicates that the 2410 is powered on.

 

 

 

Status

Yellow

The yellow Status LED blinks at a constant rate of one flash

 

 

every other second in normal operation to indicate the the

 

 

2410 software is running

 

 

 

Error

Red

The red Error LED is turned off in normal operation. If an error

 

 

occurs, the Error LED flashes a sequence that corresponds to

 

 

a certain error code, see “LED” on page 5-6.

5.5.1LED Start-Up Information

Table 5-5. LEDs are used for error indication at 2410 start-up

When the 2410 is starting, both the Status and the Error LEDs indicate possible hardware or software errors as shown in Table 5-5below:

Error type

Status LED

Error LED

Description

 

 

 

 

Hardware

Blinking

Blinking

Status and Error are blinking

 

 

 

simultaneously

 

 

 

 

Checksum

Blinking

Blinking

Status and Error are toggling

 

 

 

 

Other

On

Blinking

Unknown error