6.2.3Device Live List

Reference Manual

Rosemount 2410

300530EN, Rev BA September 2012

6.2.3Device Live List

Figure 6-1. The 2410 Tank Hub Device Live List window shows devices connected to the tankbus

The 2410 Hub Device Live List lets you view devices connected to the Tankbus. You can, for example, see the Device Id, Tag, and whether the devices are configured or not.

The Device Live List is useful when you are going to configure devices in a Raptor system, in order to verify that the required devices are connected to the Tankbus.

To view the Device Live List:

1.Start the TankMaster WinSetup program.

2.Select the Rosemount 2410 icon in the TankMaster WinSetup workspace.

3.Click the right mouse button and choose the Live List option.

The 2410 Tank Hub Device Live List window shows the following information:

Table 6-1. Device Live List

description

Item

Description

 

Device Type

Examples of supported devices: Rosemount 5900S, 2410,

 

 

2240S, 2230, 5300, 5400, 848T, and 3051S.

 

 

For unknown devices the device type number is shown.

 

Device ID (Unit ID)

A.unique code that identifies a particular device. You can

 

 

choose to view the device ID in decimal or hex format

 

 

depending on what format is supported by the device.

 

Manufact Id

Identifes the manufacturer.

 

Device No

Index used to identify devices by the FF stack.

 

FF Address

FOUNDATION Fieldbus address used for communication on the

 

 

Tankbus.

 

Handled

Bit 0 of the Live List Status input register which indicates the

 

 

current Tankbus communication status of the device.

 

Connected

“No” means that the device has been disconnected from the

 

 

Tankbus.

 

Configured

“Yes” indicates that the device is configured in the 2410 tank

 

 

database, i.e. the device is mapped to a particular tank.

 

Opened

Bit 1 of the Live List Status input register which indicates the

 

 

current Tankbus communication status of the device.

 

Auto Mode

“Yes” in normal operation. “No” indicates that the device is in

 

 

Out of Service mode.

 

Tag

A removable tag provided with the device lets you identify the

 

 

device to a physical location.