Performance Verification
4-18 2715 Spectrum Analyzer Service Manual
11. Reset the VERTICAL SCALE to 10 dB/.
12. Check that the 3rd Order IMproducts are --70 dBc.
13. Disconnect the test equipment fromthe RF INPUT.
2nd harmonic distortion performancerequirement is --66 dBc.
HTest equipmentrequired:
Signal Generator
Bandpass Filter
1. Recall default power up settings by pressing UTIL 1 1.
2. Connect the test equipment as shown in Figure 4--2.
SignalG enerator 2715 Spectrum Analyzer
BandpassF ilter
*Usemin-loss pad for
75input
*
Figure 4- 2: Harmonic Distortion TestEquipment Setup
3. Select the following settings on the Spectrum Analyzer:
REFERENCE LEVEL --40 dB m
SPAN/DIV 100 kHz
RESOLUTION BW AUTO
VID FLTR On
VERTICAL SCALE 10 dB/
4. Set the signal generator output frequencyand the Spectrum Analyzer center
frequency to the frequencyof the bandpass filter.
5. Set the signal generator output amplitudefor a full screen signal.
6. Reset the Spectrum Analyzercenter frequency to twice the input frequency.
7. Check that the 2nd harmonic is -- 66 dBc.
8. Disconnect the test equipment fromthe RF INPUT.
Harmonic Distortion