Adjustment Procedures
2715 Spectrum Analyzer Service Manual 5-65
Next you are prompted to choosebetween Calibrate Flatness, and Verify Flatness
Calibration. Choose the former to generate a table of flatnesscorrections to be
downloaded to the 2715. The data will not be automaticallydownloaded to the
2715, as with the CAL sequence. If you wish to do this, you will need to copy
the file \tekcats\flatness\2715\<sn>.ts to \tekcats\flatness\2715\<sn>.uncBe
aware that if you download afile that does not cover the full frequency or
attenuator range of the CAL sequence tests, the 2715 will not be calibrated for
CATVmeasurements at some frequency/attenuators.
Choose VerifyFlatness Calibration to check the flatness of the 2715 as corrected
by a CAL flatness run.