Adjustment Procedures
5-36 2715 Spectrum Analyzer Service Manual
Use the following procedure to perform the ServiceNormalization.
HTest equipmentrequired:
Signal Generator
Power Meter
10 dB and 1 dB Step Attenuator Assembly
Two 6 dB Attenuators
Min Loss Pad
GAIN STEP REFERENCE
Use the following procedure to normalize the gain step reference:
1. Connect the test equipment as shown in Figure 5--12.
SignalG enerator
10dB and 1 dB
StepAttenuator
Assembly
6dBPad 6dBPad
RF Output
NMale to
BNCF emale
Adapter
BNCMale to
MaleAdapter
PowerSensor
PowerMeter
50 --75
Minloss Pad
*
*Donot use for
Option50
Figure 5- 12: GainSt ep Reference TestSetup
2. Set the step attenuators to 0 dB, and set the generatorfor a 100 MHz output
and a power meter reading of --0.2 dBm for all instruments except Option 50
(0.0 dBm for Option 50 instruments).
3. Set the step attenuators to 10 dB (add 10 dB of attenuationto the 100 MHz
signal) and make a note of the P ower Meter reading. This is the actual
attenuation of the 10 dB step.
Service Normalization