Adjustment Procedures
2715 Spectrum Analyzer Service Manual 5-43
o. Set the peak carrierlevel at the top graticule line.
p. Enable the FM video mode by pressing DEMOD 3, and set the LEVEL
control for minimum flicker in the displayed picture.
q. Adjust R740 on the Sweep board for best contrastin the picture.
r. Initialize instrument settings by pressing UTIL 1 1.Ifuserdefined
power up settings exist, the in strument will recall those settings. If no
user defined power up settings exist , the instrument will recall factory