Performance Verification
2715 Spectrum Analyzer Service Manual 4-35
10dB/DIV
100US/DIV (ANALOG DISPLAY)
Figure 4- 8: IF GatingCheck Wavefor m (NTSC Illustrated)
15. Press UTIL528to turn offgating.
Performance requirement is 0dBm.
NOTE. This check applies only to Spectrum Analyzersequipped with Option 15.
HTest equipmentrequired:
Power Meter
1. Remove the 50 terminator from J101 at the S pectrum Analyzer rear panel.
2. Monitor J101 with the Power Meter.
3. Select the following settings on the SpectrumAnalyzer:
SPAN/DIV MAX SPAN
RESOLUTION BW AUTO
4. Enable MANUAL SCAN by pressing SWP/TRIG 7.
1st LO Output Level(Option 15)