Adjustment Procedures
2715 Spectrum Analyzer Service Manual 5-53
For this adjustment the 2715 is controlled by a personalcomputer using the
RS-232 or GPIB interface. The tes t equipment is controlled by GPIB. If
necessary, refer to the documentation provided wit h the GPIB interface card for
installation instructions.
The following steps describe the installation process.
HInstall the test software
HRun the system configuration program(CONF IG)
HConnect the personal computerto control the test instrument
HRun the test equipment characterizationprogram (CHAR)
NOTE. These steps are requiredfor initial installation only. After the initial
installation, it is not necessary to run CONFIG or CHAR as long as the system
remains intact.
Install the Flatness TestSoftware. The F latness Test software is contained on a set
of floppy disks (271X Flatness Tests)supplied with this manual. Use the
following procedure to install the software onto the personalcomputer:
1. Install disk 1 in drive A of the personalcomputer.
2. Set the default drive to A by typing:
A:<Return>
3. Type the following commandto install the flatness software on your hard
disk:
Install<Return>
A prompt will instruct you when to switch disks (for5. 25 disks only).
4. Reboot your PC to incorporate the changes to the autoexec.bat file.
NOTE. INSTALL.BAT is a batch file that createsa directory on your hard disk
(called TEKCATS) and several subdirectories. It then copies the Flatness Test
software into these directories.
Computer Setup andSoftware Installation