Performance Verification
2715 Spectrum Analyzer Service Manual 4-21
Performance requirements are:
Pin 5: Video Out, Pin 2: Sweep Gate, Pin 4: Sweep Output
HTest equipmentrequired:
Test Oscilloscope
1. Recall default power up settings by pressing UTIL 1 1.
2. Select the following settings on the SpectrumAnalyzer:
FREQUENCY 100 MHz
REFERENCE LEVEL --30 dB m
SPAN/DIV 1 MHz
RESOLUTION BW AUTO
VID FLTR Off
VERTICAL SCALE 10 dB/
DISPLAY STORAGE OFF
3. Enable the calibrator by pressing INPUT 9.
4. Disable frequency correctionsby pressing UTIL435.
5. Monitor Pin 5 of J103 (Figure 4-- 4)with the test oscilloscope. Set the test
oscilloscope sweep rateto 10 ms/div and Volts/div to 0.5.
J103 AccessoryConnector