Adjustment Procedures
2715 Spectrum Analyzer Service Manual 5-47
Adjust Video Detector. Adjust L11 as follows:
1. Connect the TSG 95 VideoOutput to the TVM450 Video Input. Connect the
TVM450s --20 dB RF TES T Output to the2715 R F INPUT.
2. Set the TSG 95 to produce a 5-Step Grayscaleand set it to the desired video
format (NTSC or PAL).
3. Set the TVM450 output to Channel 4 and the RF level control full counter-
clockwise.
NOTE. Be sure the OFFSET switch on the rearpanel of the TVM450 is set to OFF.
4. Check that the VIDEO MODULATIONand VIDEO/AUDIO RATIO
controls on the TVM450 are both set to mid range(detent) and that the
AUDIO DEVIATIONcontrol is set full counterclockwise.
5. Select the following settings on the 2715:
CENTER FREQUENCY 67. 25 MHz
SPAN/DIV 2 MHz
Calibrator Signal OFF
6. Set the signal (visual carrier) to the top graticulewith the REF LEVEL and
controls.
7. Set SPAN to ZERO.
8. Activate the VD DAC TO VIS CARR function by pressing UTIL526.
9. Select the following settings on the test oscilloscope:
VerticalMode CH1
CH 1 Volts/Div. 500 mV
CH 1 Coupling DC
Time/Div. 10s
Trigger Mode Auto
Trigger Source CH 1
Trigger Coupling DC
10. Connect the oscilloscope channel 1 probe to the VideoDemod VIDEO TP.
11. Tune L11 until the test oscilloscopedisplays a video signal as illustrated in
Figure 5--18.
NOTE. It will probably be necessaryto adjust the test oscilloscope trigger level
control to obtain a stable display.