Performance Verification
2715 Spectrum Analyzer Service Manual 4-29
5. Press B--Save A on the test spectrum analyzer.A straight line will be at
midscreen. This is the reference t race.
6. Connect the VSWR Bridge to the RF INPUT and set the REFERENCE
LEVEL to --20 dBm.
7. A frequency response waveformwill be displayed on the Test Spectrum
Analyzer.
8. Check that this waveform is 1.5 divisions down from the referencetrace
across the entire band.
Use the following procedureto verify operation of the Signal Track feature.
1. Recall default power up settings by pressing UTIL 1 1.
2. Select the following settings on the SpectrumAnalyzer:
FREQUENCY 100 MHZ
REFERENCE LEVEL --20 dB m
SPAN/DIV 100 kHz
RESOLUTION BW AUTO
VID FLTR Off
VERTICAL SCALE 5 dB/
3. Enable the calibrator by pressing INPUT 9.
4. Press CTR-MEAS/TRKG twice.
5. Check that the letters TRK G appear at center screen.
6. Tune the signal away from the centerof the screen.
7. Check that the signal return s to the center of the screen.
Signal Track