142 Chapter 4

ESA Functional Tests

Frequency Response (Flatness)

ESA Functional Tests
Frequency Response (Flatness)
Test Limits
Test Description

The output of the synthesized sweeper is fed through a power splitter to a power sensor

and the analyzer. The synthesized sweepers power level is ad justed at 50 MHz to place the

displayed signal at the analyzer center horizontal graticule line. Measurements are made

at various points depending on the model being tested. The signal source amplitude is

measured with a power meter to eliminate errors due to source flatness. The power meter

is zeroed and calibrated before starting the measurement.

ESA Model Frequency Minimum
(dB) Maximum
(dB)
E4411B & E4401B 9 kHz to 1.5 GHz 1.5 1.5
E4402B & E4403B 9 kHz to 3 GHz 1.5 1.5
E4404B 9 kHz to 3 GHz 1.5 1.5
3.01 GHz to 6.7 GHz 3.0 3.0
E4405B 9 kHz to 3 GHz 1.5 1.5
3.01 GHz to 6.7 GHz 3.0 3.0
6.71 GHz to 13.2 GHz 3.5 3.5
E4407B & E4408B 9 kHz to 3 GHz 1.5 1.5
3.01 GHz to 6.7 GHz 3.0 3.0
6.71 GHz to 13.2 GHz 3.5 3.5
13.21 GHz to 25 GHz 4.0 4.0
25 GHz to 26.5 GHz 4.5 4.5
Item Critical Specifications
(for this test) Recommended
Agilent Model
Signal Sources
Synthesized
Sweeper 8340A/B or 836xx Series
Adapters
Type-N (f) to BNC (f) 1250-1474
APC 3.5 (f) to APC
3.5 (f) 5061-5311
BNC (f) to SMA (m) 1250-1237