Power Meter Operation
Recorder Outputs
Leveling a Source Output
The Recorder output can be used to level an output from a source using external leveling. The following procedure explains how to do this:
1. Select the Recorder Output A or B, depending on which channel
the source is connected to, by pressing System , More ,
Inputs
Recorder Output , Channel A or Channel B .
2.The highest power you are going to measure is used to determine the value which you should set for the Recorder Output maximum
setting. For example, if you are measuring a power less than 1 mW and greater than 100 μW, then set the recorder maximum value to 1 mW.
50 dBm (100 W)
40 dBm (10 W)
30 dBm (1 W)
20 dBm (100 mW)
10 dBm (10 mW)
0 dBm (1 mW)
To set the maximum value press Max Power and enter the appropriate value.
3. Press Min Power and enter 0 W.
4. Press
Output Off On to “On”.
HP E4419B User’s Guide |