Power Meter Operation
Making Measurements with the HP 8480 Series Power Sensors
•Press A Ref CF . Use the ,
,
and
hardkeys to enter 99.8. Press % .
•Connect the power sensor to the POWER REF output.
•Press Cal A .
• Press | Frequency | , A Cal Fac . Use the | , | , | and |
Cal Fac |
hardkeys to enter 97.8. Press % .
•Connect the power sensor to the signal to be measured.
•The measurement result is now displayed.
HP E4419B User’s Guide |