
Power Meter Operation
Zeroing and Calibrating the Power Meter
If you calibrate the sensor before zeroing it, the message changes to “Please Zero ChA” or “Please Zero ChB” depending on which channel the sensor is connected to.
The Zero/Cal Lockout facility can be enabled/disabled through either the System Inputs menu or the Zero Cal menu as follows:
Press System , More , More , Inputs
Must Cal
Off or
On.
Press ZerCalo , More , Must Cal
Off or
On.
| Calibrating the Power Meter |
| Calibration sets the gain of the power meter using a 50 MHz 1 mW |
| calibrator as a traceable power reference. The power meter’s POWER REF |
| output or a suitable external reference is used as the signal source for |
| calibration. An essential part of calibrating is setting the correct reference |
| calibration factor for the power sensor you are using. The HP 8480 series |
| power sensors require you to set the reference calibration factor. The |
| HP |
| automatically. During calibration the wait symbol is displayed. Offset, |
| relative and duty cycle settings are ignored during calibration. |
|
|
Note | During calibration the power meter automatically switches the |
| power reference calibrator on (if it is not already on), then after |
| calibration it switches it to the state it was in prior to the |
| calibration. |
|
|
Calibration Procedure Using HP E-Series Power Sensors
The following procedure describes how you calibrate the power meter with an HP
HP E4419B User’s Guide |