HP Saw E4419B manual Data corrupt or stalePlease zero and calibrate Channel a

Models: Saw E4419B

1 204
Download 204 pages 1024 b
Page 174
Image 174

Error Messages

Error Messages

-230

Data corrupt or stale;Please zero and calibrate Channel A

 

When CAL[12]:RCAL is set to ON and the sensor currently connected

 

to channel A has not been zeroed and calibrated, then any command

 

which would normally return a measurement result (for example

 

FETC?, READ?, or MEAS?) will generate this error message.

-230

Data corrupt or stale;Please zero and calibrate Channel B

 

When CAL[12]:RCAL is set to ON and the sensor currently connected

 

to channel B has not been zeroed and calibrated, then any command

 

which would normally return a measurement result (for example

 

FETC?, READ?, or MEAS?) will generate this error message.

-230

Data corrupt or stale;Please zero Channel A

 

When CAL[12]:RCAL is set to ON and the sensor currently connected

 

to channel A has not been zeroed, then any command which would

 

normally return a measurement result (for example FETC?, READ?, or

 

MEAS?) will generate this error message.

-230

Data corrupt or stale;Please zero Channel B

 

When CAL[12]:RCAL is set to ON and the sensor currently connected

 

to channel B has not been zeroed, then any command which would

 

normally return a measurement result (for example FETC?, READ?, or

 

MEAS?) will generate this error message

-230

Data corrupt or stale;Please calibrate Channel A

 

When CAL[12]:RCAL is set to ON and the sensor currently connected

 

to channel B has not been calibrated, then any command which would

 

normally return a measurement result (for example FETC?, READ?, or

 

MEAS?) will generate this error message

-230

Data corrupt or stale;Please calibrate Channel B

 

When CAL[12]:RCAL is set to ON and the sensor currently connected

 

to channel B has not been calibrated, then any command which would

 

normally return a measurement result (for example FETC?, READ?, or

 

MEAS?) will generate this error message

-231

Data questionable;CAL ERROR ChA

 

Power meter calibration failed on channel A. The most likely cause is

 

attempting to calibrate without applying a 1 mW power to the power

 

sensor.

4-8

HP E4419B User’s Guide

Page 174
Image 174
HP Saw E4419B manual Data corrupt or stalePlease zero and calibrate Channel a, Data corrupt or stalePlease zero Channel a