Error Messages
Error Messages
System error;Ch B Dty Cyc may impair accuracy with ECP | |
| sensor |
| This indicates that the sensor connected to channel B is for use with |
| CW signals only. |
System error;Detector EEPROM Read Failed - critical data not | |
| found or unreadable |
| This indicates a failure with your HP |
| your power sensor manual for details on returning it for repair. |
System error;Detector EEPROM Read Completed OK but | |
| optional data block(s) not found or unreadable |
| This indicates a failure with your HP |
| your power sensor manual for details on returning it for repair. |
System error;Detector EEPROM Read Failed - unknown | |
| EEPROM table format |
| This indicates a failure with your HP |
| your power sensor manual for details on returning it for repair. |
System error;Detector EEPROM < > data not found or | |
| unreadable |
| Where < > refers to the sensor data block covered, for example, |
| Linearity, Temp - Comp (temperature compensation). |
| This indicates a failure with your HP |
| your power sensor manual for details on returning it for repair. |
System error;Option 001 Battery charger fault | |
| The power meter is connected to an ac power source, the battery is not |
| fully charged and it is not charging. |
System error;Sensors connected to both front and rear inputs. | |
| You cannot connect two power sensors to the one channel input. In this |
| instance the power meter detects power sensors connected to both it’s |
| front and rear channel inputs. |
Out of memory | |
| The power meter required more memory than was available to run an |
| internal operation. |
HP E4419B User’s Guide |