![](/images/new-backgrounds/1330617/330617373x1.webp)
Specifications
Power Meter Supplemental Characteristics
Settling Time
0 to 99% settled readings over the
For HP 8480 series power sensors
Manual filter, 10 dB decreasing power step:
Table 5-4: Settling Time
Number of |
| 1 | 2 | 4 | 8 | 16 | 32 | 64 | 128 | 256 | 512 | 1024 |
Averages |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
Settling Time (s) |
| 0.15 | 0.2 | 0.3 | 0.5 | 1.1 | 1.9 | 3.4 | 6.6 | 13 | 27 | 57 |
(Normal Mode) |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
Response Time (s) |
| 0.15 | 0.18 | 0.22 | 0.35 | 0.55 | 1.1 | 1.9 | 3.5 | 6.9 | 14.5 | 33 |
(x2 mode) |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
Auto filter, default resolution, 10 dB decreasing power step, normal and x2 speed modes:
| Normal Mode |
|
| x2 Mode |
|
| ||
| Maximum dBm |
|
| Maximum dBm |
| |||
| 150 ms | 20 dB |
|
| 150 ms | 20 dB |
| |
Typical |
|
| Power | Typical |
|
| Power | |
Settling | 200 ms | 10 dB | Sensor | Settling | 180 ms | 10 dB | Sensor | |
Times | Dynamic | Times | Dynamic | |||||
|
|
|
| |||||
|
|
| Range |
|
|
| Range | |
| 500 ms | 10 dB |
|
| 350 ms | 10 dB |
| |
| 6.6 s | 10 dB |
|
| 3.5 s | 10 dB |
| |
| Minimum dBm |
|
| Minimum dBm |
|
HP E4419B User’s Guide |