Error Messages

Error Messages

-330

Self-test Failed;

 

The -330, “Self-test Failed” errors indicate that you have a problem

 

with your power meter. Refer to “Contacting Hewlett-Packard”, on

 

page 2-81 for details of what to do with your faulty power meter.

-330

Self-test Failed;Measurement Channel A Fault

 

Refer to “Measurement Assemblies”, on page 2-77 if you require a

 

description of the Measurement Assembly test.

-330

Self-test Failed;Measurement Channel B Fault

 

Refer to “Measurement Assemblies”, on page 2-77 if you require a

 

description of the Measurement Assembly test.

-330

Self-test Failed;Option 001 Battery requires replacement

 

The option 001 battery is not charging to a satisfactory level and should

 

be replaced.

-330

Self-test Failed;RAM Battery Fault

 

Refer to “RAM Battery”, on page 2-77 if you require a description of the

 

battery test.

-330

Self-test Failed;Calibrator Fault

 

Refer to “Calibrator”, on page 2-78 if you require a description of the

 

calibrator test.

-330

Self-test Failed;ROM Check Failed

 

Refer to “ROM Checksum”, on page 2-77 if you require a description of

 

the ROM Checksum test.

-330

Self-test Failed;RAM Check Failed

 

Refer to “RAM”, on page 2-77 if you require a description of the RAM

 

test.

-330

Self-test Failed;Display Assy. Fault

 

Refer to “Display”, on page 2-78 if you require a description of the

 

Display test.

-330

Self-test Failed;Confidence Check Fault ChA

 

Refer to “Confidence Check”, on page 2-75 if you require a description of

 

this test.

-330

Self-test Failed;Confidence Check Fault ChB

 

Refer to “Confidence Check”, on page 2-75 if you require a description of

 

this test.