Menu Reference
Front Panel Menu Reference "Frequency/Cal Fac" Menu
•
A Cal Fac
Press this softkey to enter the measurement calibration factor for channel A in the range of 1% to 150%. The default value is 100%.
Use the , , and hardkeys to change the calibration factor. To confirm your choice press % . Entering a calibration factor using this softkey is only possible when you are using an HP 8480 series power sensor.
•B Freq
Press this softkey to enter the frequency of the signal to be measured on channel B, in the range of 0.001 MHz to
999.999 GHz. The default value is 50 MHz. Use the , ,
and hardkeys to change the frequency. To confirm your choice press the appropriate frequency unit. Entering a frequency using this softkey is only possible when you have a sensor calibration table selected or you are using an HP
•
B Cal Fac
Press this softkey to enter the measurement calibration factor for channel B in the range of 1% to 150%. The default value is 100%.
Use the , , and hardkeys to change the calibration factor. To confirm your choice press % . Entering a calibration factor using this softkey is only possible when you are using an HP 8480 series power sensor.
HP E4419B User’s Guide |