Power Meter Operation
Self Test
Self Test
The power meter has three distinct self test modes:
•power on self test, which occurs automatically when you turn on the power meter.
•confidence check, which is accessed via the front panel and requires your assistance in order to verify the POWER REF and measurement path accuracy.
•troubleshooting mode, which is accessed via the front panel or remotely. The front panel softkey menu allows you to run individual tests, whereas the remote command runs a complete series of tests as listed in “Remote Testing”, on page
Power On Self Test
The power on self test is performed automatically when the power meter is turned on and takes approximately 10 seconds to complete. The power on self test consists of the following tests:
•RAM Battery
•Calibrator
•Measurement Assemblies
•Fan
•Serial Interface
•Option 001 Battery
Refer to “Test Descriptions”, on page
As the power on self test takes place, the message “Testing...” appears next to the name of the test which is being performed. When each test is complete, the message “Testing...” is replaced by either the message “Passed” or “Failed”. The option 001 Battery test can also produce the message “Not Present”. If a failure occurs the message
examined in the “Errors” screen by pressing , More , Error List .
HP E4419B User’s Guide |