![](/images/new-backgrounds/1330617/330617355x1.webp)
Error Messages
Error Messages
| |
| Refer to “Serial Interface”, on page |
| this test. |
Queue overflow | |
| The error queue is full and another error has occurred which could not |
| be recorded. |
Parity error in program | |
| The serial port receiver has detected a parity error and consequently, |
| data integrity cannot be guaranteed. |
Framing error in program | |
| The serial port receiver has detected a framing error and consequently, |
| data integrity cannot be guaranteed. |
Input buffer overrun | |
| The serial port receiver has been overrun and consequently, data has |
| been lost. |
Query INTERRUPTED | |
| A command was received which sends data to the output buffer, but the |
| output buffer contained data from a previous command (the previous |
| data is not overwritten). The output buffer is cleared when power has |
| been off, or after *RST (reset) command has been executed. |
Query UNTERMINATED | |
| The power meter was addressed to talk (that is, to send data over the |
| interface) but a command has not been received which sends data to the |
| output buffer. For example you may have executed a CONFigure |
| command (which does not generate data) and then attempted to read |
| data from the remote interface. |
Query DEADLOCKED | |
| A command was received which generates too much data to fit in the |
| output buffer and the input buffer is also full. Command execution |
| continues but data is lost. |
Query UNTERMINATED after indefinite response | |
| The *IDN? command must be the last query command within a |
| command string. |
HP E4419B User’s Guide |