Error Messages

Error Messages

-330

Self-test Failed;Serial Interface Fault

 

Refer to “Serial Interface”, on page 2-77 if you require a description of

 

this test.

-350

Queue overflow

 

The error queue is full and another error has occurred which could not

 

be recorded.

-361

Parity error in program

 

The serial port receiver has detected a parity error and consequently,

 

data integrity cannot be guaranteed.

-362

Framing error in program

 

The serial port receiver has detected a framing error and consequently,

 

data integrity cannot be guaranteed.

-363

Input buffer overrun

 

The serial port receiver has been overrun and consequently, data has

 

been lost.

-410

Query INTERRUPTED

 

A command was received which sends data to the output buffer, but the

 

output buffer contained data from a previous command (the previous

 

data is not overwritten). The output buffer is cleared when power has

 

been off, or after *RST (reset) command has been executed.

-420

Query UNTERMINATED

 

The power meter was addressed to talk (that is, to send data over the

 

interface) but a command has not been received which sends data to the

 

output buffer. For example you may have executed a CONFigure

 

command (which does not generate data) and then attempted to read

 

data from the remote interface.

-430

Query DEADLOCKED

 

A command was received which generates too much data to fit in the

 

output buffer and the input buffer is also full. Command execution

 

continues but data is lost.

-440

Query UNTERMINATED after indefinite response

 

The *IDN? command must be the last query command within a

 

command string.

4-12

HP E4419B User’s Guide

Page 178
Image 178
HP Saw E4419B Self-test FailedSerial Interface Fault, Queue overflow, Parity error in program, Framing error in program