Power Meter Operation

Zeroing and Calibrating the Power Meter

Zeroing and Calibrating Using TTL Inputs

You can use the TTL inputs on the rear panel Rmt I/O port to initiate zero and calibration cycles on the power meter. The connector is an RJ-45 series shielded modular jack with the TTL input pins connected as shown in Figure 2-3.

Figure 2-3: Rmt I/O Port TTL Inputs

1 2 3 4 5 6 7 8

Pin Number

Connection

 

 

1

none

 

 

2

Ground

 

 

3

Upper Window TTL Output

 

 

4

Lower Window TTL Output

 

 

5

TTL Input 1

 

 

6

TTL Input 2

 

 

7

Ground

 

 

8

Ground

 

 

The TTL inputs are active low and control the zero and calibration functions as shown in Table 2-2

Table 2-2: TTL Inputs Control Logic

Input 1

Input 2

Operation

 

 

 

1

1

None

 

 

 

1

0

CAL A

 

 

 

0

1

ZERO BOTH

 

 

 

0

0

CAL B

 

 

 

HP E4419B User’s Guide

2-13

Page 53
Image 53
HP Saw E4419B manual Zeroing and Calibrating Using TTL Inputs, TTL Inputs Control Logic Operation