Power Meter Operation
Setting Measurement Limits
To set the TTL Output:
1.Press SetupMeas , Limits , TTL Output . The current settings for the TTL Output are displayed under the TTL Output, Limits and Fail O/P softkeys.
2.To select whether the TTL output reflects an overlimit condition, under limit condition or both, press Limits . Use any of the four arrow keys to make a selection between OVER , UNDER , and EITHER on the pop up menu that appears.
3.To select whether a high or low level TTL output represents a limits failure, use the Fail O/P softkey to toggle between High and Low
TTL Outputs can be disabled and
TTL Output Off On .
Checking for Limit Failures
Limit failures are displayed in the appropriate field in the measurement window on the power meter’s display as shown in Figure
Note | The same limit fail field is used by both channel and window |
| limits. |
|
|
HP E4419B User’s Guide |