Error Messages

Error Messages

-148

Character data not allowed

 

A discrete parameter was received but a character string or a numeric

 

parameter was expected. Check the list of parameters to verify that you

 

have used a valid parameter type.

 

For example, MEM:CLE CUSTOM_1.

-151

Invalid string data

 

An invalid string was received. Check to see if you have enclosed the

 

character string in single or double quotes.

 

For example, MEM:CLE “CUSTOM_1.

-158

String data not allowed

 

A character string was received but is not allowed for the command.

 

Check the list of parameters to verify that you have used a valid

 

parameter type.

 

For example, LIM:STAT ‘ON’.

-161

Invalid block data

 

A block data element was expected but was invalid for some reason. For

 

example, *DDT #15FET. The 5 in the string indicates that 5 characters

 

should follow, whereas in this example there are only 3.

-168

Block data not allowed

 

A legal block data element was encountered but not allowed by the

 

power meter at this point.

 

For example SYST:LANG #15FETC?.

-178

Expression data not allowed

 

A legal expression data was encountered but not allowed by the power

 

meter at this point.

 

For example SYST:LANG (5+2).

-211

Trigger ignored

 

Indicates that <GET> or *TRG, or TRIG:IMM was received and

 

recognized by the device but was ignored because the power meter was

 

not in the wait for trigger state.

-213

Init ignored

 

Indicates that a request for a measurement initiation was ignored as

 

the power meter was already initiated.

 

For example, INIT:CONT ON

 

INIT.