HP Saw E4419B manual Off

Models: Saw E4419B

1 204
Download 204 pages 1024 b
Page 163
Image 163

Menu Reference

Front Panel Menu Reference "Zero/Cal" Menu

Cal B

Press this softkey to calibrate channel B with the connected power sensor. The POWER REF output is used as the signal source for calibration and is automatically switched on during this procedure.

B Ref CF

Press this softkey to enter a reference calibration factor for channel B. A value in the range of 1% to 150% can be entered. The default value is obtained from the sensor calibration table

if one is selected, otherwise it is 100%. Use the , ,

and hardkeys to change the value. To confirm your choice press % . This softkey cannot be selected when you are using an HP E-series power sensor power sensor.

Power Ref Off On

 

Press this softkey to switch the POWER REF output on or off. This

 

output is used as the signal source for calibration. The default is

 

Off”.

 

 

Note

During calibration the power meter automatically switches the

 

power reference oscillator on (if it is not already on), then after

 

calibration it switches it to the state it was in prior to the

 

calibration.

 

 

Must Cal Off On

Press this softkey to switch the Zero/Cal Lockout facility On or Off. When Zero/Cal Lockout is On, the power meter is prevented from making a measurement until the sensor connected has been zeroed and calibrated. This softkey is coupled to the equivalent softkey in the System Inputs menu.

TTL Inputs Off On

Press this softkey to disable/enable (Off/On) the ZERO and CAL rear panel TTL inputs.

HP E4419B User’s Guide

3-35

Page 163
Image 163
HP Saw E4419B manual Off