Menu Reference

Front Panel Menu Reference "System/Inputs" Menu

Echo

Press this softkey to toggle Echo “On” or “Off”. When Echo is enabled, received characters are transmitted back to the sender.

Tables

Press this softkey to access a menu which allows you to select between sensor calibration tables and frequency dependent offset tables.

Sensor Cal Tables

Press this softkey to access a menu and screen which allows you to select and edit sensor calibration tables. Use the

and hardkeys to scroll through the displayed sensor calibration tables. Sensor calibration tables are only required for the HP 8480 series power sensors.

Edit Table

Press this softkey to access a menu and screen which allows you to modify, add and remove frequency, calibration factors and offsets from the currently selected table and modify the

table name. Use the , , and hardkeys to move between the name, frequency, calibration factors and offsets. Press Change, Insert or Delete as follows:

Change

Press this softkey to modify the highlighted parameter, which is either a frequency, calibration factor, offset or table name. A pop up window displays the parameter. Use

the , , and hardkeys to change the parameter value. To confirm your choice press Enter .

Insert

Press this softkey to insert a new table entry. You are prompted to enter the frequency and calibration factor or offset. The entry is inserted in ascending frequency order.

Delete

Press this softkey to delete a table entry. If you delete the frequency the corresponding calibration factor or offset is also removed and vice versa.

Done

Press this softkey to return to the measurement screen. The softkeys from the first level of the “System/Inputs” menu are displayed.

3-24

HP E4419B User’s Guide