Power Meter Operation

Presetting the Power Meter

Presetting the Power Meter

This section details the power meter’s preset conditions.

The HP-IB address, the data stored in the sensor calibration tables and the zeroing and calibration data are not affected by a preset. The calibration table selected is not affected.

Preset Conditions

The number of windows displayed is two.

dBm/W

Measurement units used is set to dBm.

Frequency

Cal Fac

Meas Setup

AFreq is set to 50 MHz.

ACal Fac is set to 100%.

BFreq is set to 50 MHz.

B Cal Fac is set to 100%.

Input Select is set to Channel A in the upper window and Channel B in the lower window.

Display Format is set to digital for both windows.

Anlg Mtr Scaling is set to 20.000 dBm ( Max ) and -70.000 dBm ( Min ).

Resolution 1 2 3 4 is set to “3”.

Limits Off/On is set to “Off

Limits is set to 90.000 dbm ( Max ) and -90.000 dBm ( Min ). TTL Output Off/On is set to “Off”.

Limits OVER UNDER EITHER is set to “OVER”. Fail O/P High Low is set to “Low

Rel

Offset

Rel Off/On is set to “Off”.

Offset Off/On is set to “Off”.

2-70

HP E4419B User’s Guide

Page 110