Error Messages

Introduction

Introduction

This chapter contains information about the power meter’s error messages. It explains how to read the power meter’s error queue and lists all of the power meter’s error messages and their probable causes.

When there is a hardware related problem, for example, a power sensor overload, the error message is displayed on the status line at the top of the display. In addition, the errors are also written to the error queue. If there are any errors in the error queue the front panel error annunciator is displayed as shown in Figure 4-1.

Other errors can also be generated when the power meter is being operated over the remote interface. These errors also display the error annunciator and are written to the error queue.

Figure 4-1: Error Annunciator Position

Error annunciator

To read the error queue from the front panel press:

System

, More , Error List then use Next to scroll through

Inputs

each error message.

To read the error queue from the remote interface use:

the SYSTem:ERRor? command.

Error queue messages have the following format:

Error Number

,

Error

;

Device

 

 

Description

 

Dependent Info

 

For example, -330, “Self-test Failed;Battery Fault”.

4-2

HP E4419B User’s Guide

Page 168
Image 168
HP Saw E4419B manual Error Annunciator Position