Error Messages
Introduction
Introduction
This chapter contains information about the power meter’s error messages. It explains how to read the power meter’s error queue and lists all of the power meter’s error messages and their probable causes.
When there is a hardware related problem, for example, a power sensor overload, the error message is displayed on the status line at the top of the display. In addition, the errors are also written to the error queue. If there are any errors in the error queue the front panel error annunciator is displayed as shown in Figure
Other errors can also be generated when the power meter is being operated over the remote interface. These errors also display the error annunciator and are written to the error queue.
Figure 4-1: Error Annunciator Position
Error annunciator
To read the error queue from the front panel press:
• | System | , More , Error List then use Next to scroll through |
Inputs |
each error message.
To read the error queue from the remote interface use:
•the SYSTem:ERRor? command.
Error queue messages have the following format:
Error Number
, | “ | Error | ; | Device | ” |
|
| Description |
| Dependent Info |
|
For example,
HP E4419B User’s Guide |