Power Meter Operation
Zeroing and Calibrating the Power Meter
Note | The following procedure details the calibration for channel A. To | |||||||
| calibrate channel B use the same procedure using the equivalent | |||||||
| channel B softkeys. | |||||||
|
|
|
|
|
|
|
|
|
| 1. | Press |
| . |
|
| ||
| Zero | |||||||
| Cal | |||||||
| 2. | Connect the power sensor to the POWER REF output. | ||||||
| 3. | Press |
|
| , |
| to calibrate the power meter. During | |
| Cal |
| Cal A | |||||
|
| calibration the wait symbol is displayed. (The power meter | ||||||
|
| automatically turns on the POWER REF output.) |
Example
To calibrate both channels of the power meter with HP
• Press Zero
Cal
,
Cal .
•Connect the channel A power sensor to the POWER REF output.
•Press
Cal A then wait till the calibration is completed.
•Remove the channel A power sensor from the POWER REF output.
•Connect the channel B power sensor to the POWER REF output.
•Press
Cal B then wait till the calibration is completed.
| Calibration Procedure using HP 8480 Series Power Sensors |
| The following procedure describes how you calibrate the power meter with |
| the HP 8480 series power sensors. |
|
|
Note | V8486A and W8486A sensors |
| For most 8480 series sensors the correct (A type or D type) |
| linearity correction table is automatically selected. However, for |
| the V8486A and W8486A sensors the automatic selection must be |
| overriden and the D type correction selected. Subsequent |
| connection of another A type sensor will result in a warning |
| message stating that “Linearity Override May be Required”. |
To select the linearity type to be applied:
Press | System | , Tables, A Linearity ATyp or |
Inputs |
B Linearity ATyp or DTyp .
DTyp , or
HP E4419B User’s Guide |