Menu Reference
Front Panel Menu Reference "System/Inputs" Menu
■
Next
Press this softkey to move to the next error in the error queue. The displayed error message is individually cleared each time Next is selected.
■Done
Press this softkey to return to the “System/Inputs” menu.
•
Recorder Output
Press this key to access a menu which allows you to select either Channel A or Channel B recorder output for setup modification.Refer to “Recorder Outputs”, on page
■
Output Off On
Press this softkey to switch the rear panel Recorder Output on or off. The Recorder Output produces a dc voltage that corresponds to the selected input channel power level in Watts.
■
Max Power
Press this softkey to enter the input power level that you want to represent the 1 Vdc maximum output voltage of the Recorder Output.
■
Min Power
•
Press this softkey to enter the input power level that you want to represent the 0 Vdc minimum output voltage of the Recorder Output.
Service
Press this softkey to access a menu which allows you to test and service your power meter.
■Self Test
Press this softkey to access the power meter’s self test menu.
◆Instrument Self Test
Press this softkey to run a series of tests on the power meter. Refer to “Instrument Self Test”, on page
◆
Confidence Check
Press this softkey to verify that the power meter makes an accurate measurement of it’s 1 mW POWER REF output.