Power Meter Operation
Zeroing and Calibrating the Power Meter
Table
Sensor | Connection Requirements | |
Model | ||
| ||
|
| |
|
| |
HP 8481A | These power sensors connect directly to the reference calibrator. | |
HP 8481H |
| |
HP 8482A |
| |
HP 8482H |
| |
|
| |
HP 8481D | Prior to the power meter being calibrated an HP 11708A 30 dB | |
HP 8484A | reference attenuator should be connected between the power | |
| sensor and the reference calibrator. This attenuator must be | |
| removed from the power sensor input prior to making | |
| measurements. | |
|
| |
HP 8483A | This power sensor requires a 75 Ω (f) to 50 Ω (m) | |
| ||
| must be removed from the power sensor input prior to making | |
| measurements. | |
|
| |
HP R8486A | The waveguide power sensors have two connectors. The | |
HP Q8486A | connector is the one which is used to calibrate the power meter. | |
HP V8486A |
| |
HP W8486A |
| |
HP R8486D |
| |
HP Q8486D |
| |
|
| |
HP 8481B | These power senors are configured with an attenuator. Prior to | |
HP 8482B | the power meter being calibrated, this attenuator must be | |
| removed. The attenuator must be reconnected prior to making | |
| measurements. | |
|
| |
HP 8485A | This power sensor requires an APC 3.5 (f) to 50 Ω (m) | |
| adapter | |
|
| |
HP 8485D | Prior to the power meter being zeroed and calibrated an | |
| HP 11708A 30 dB reference attenuator and an APC 3.5 (f) to | |
| 50 Ω (m) | |
| between the power sensor and the reference calibrator. This | |
| attenuator must be removed from the power sensor input prior to | |
| making measurements. | |
|
| |
HP 8487A | This power sensor requires an APC 2.4 (f) to 50 Ω (m) | |
| adapter | |
|
| |
HP 8487D | Prior to the power meter being zeroed and calibrated an | |
| HP 11708A 30 dB reference attenuator and an APC 2.4 (f) to | |
| 50 Ω (m) | |
| between the power sensor and the reference calibrator. This | |
| attenuator must be removed from the power sensor input prior to | |
| making measurements. | |
|
|
HP E4419B User’s Guide |