Power Meter Operation

Setting Averaging

Step Detection

To reduce the filter settling time after a significant step in the measured power the filter can be set to re-initialise upon detection of a step increase/decrease in the measured power. Step detection can be set in both manual and automatic filter modes.

To set step detection for channel A:

1. Press System

Inputs

,

A Input Settings , More .

2. Press the

ChA Filter softkey to access the filter menu.

3.Use the Step Det Off On softkey to enable /disable step detection.

For channel B use the steps above except use ChB Filter .

B Input Settings and

HP E4419B User’s Guide

2-43

Page 83
Image 83