Power Meter Operation

Configuring the Remote Interface

Configuring the Remote Interface

This section explains how to configure the power meter for remote control using either the HP-IB parallel interface or the RS232/422 serial interface. For further information on using the power meter remotely, refer to the HP E4418B/E4419B Programming Guide.

HP-IB

To select HP-IB as the remote interface press

System

,

Inputs

Remote Interface, Select Interface, HPIB (IEEE 488).

HP-IB Address

Each device on the HP-IB (IEEE-488) interface must have a unique address. You can set the power meter’s address to any value between 0 and 30. The address is set to 13 when the power meter is shipped from the factory.

The address is stored in non-volatile memory, and does not change when the power is switched off, or after a remote interface reset.

Your HP-IB bus controller has its own address. Avoid using the bus controller’s address for any instrument on the interface bus. Hewlett-Packard controllers generally use address 21.

To set the HP-IB address from the front panel:

1. Press System , Remote Interface, Configure Interface ,

Inputs

HPIB . The current setting of the HP-IB address is displayed under the HP-IB Addr softkey.

2.To change this setting press HP-IB Addr . The power meter displays the address in a pop up window. Modify this address (see below) as desired.

Use or to modify the digit on which the cursor is currently positioned.

Use or to move to other digits.

3.To confirm your choice press Enter .

To set the HP-IB address from the remote interface use the:

SYSTem:COMMunicate:GPIB:ADDRess command.

HP E4419B User’s Guide

2-59