Error Messages
Error Messages
Trigger deadlock | |
| TRIG:SOUR was set to HOLD or BUS and a READ? or MEASure? was |
| attempted, expecting TRIG:SOUR to be set to IMMediate. |
Parameter error;Frequency list must be in ascending order. | |
| Indicates that the frequencies entered using the |
| MEMory:TABLe:FREQuency command are not in ascending order. |
Settings conflict | |
| This command occurs under a variety of conflicting conditions. The |
| following list gives a few examples of where this error may occur: |
| •If the READ? parameters do not match the current settings. |
| •If you are in fast mode and attempting to switch on for example, |
| averaging, duty cycle or limits. |
| •Trying to clear a sensor calibration table when none is selected. |
Settings conflict;DTR/DSR not available on RS422 | |
| DTR/DSR is only available on the RS232 interface. |
Data out of range | |
| A numeric parameter value is outside the valid range for the command. |
| For example, SENS:FREQ 2KHZ. |
Illegal parameter value | |
| A discrete parameter was received which was not a valid choice for the |
| command. You may have used an invalid parameter choice. |
| For example, TRIG:SOUR EXT. |
Lists not same length | |
| This occurs when SENSe:CORRection:CSET[1]CSET2:STATe is set |
| to ON and the frequency and calibration/offset lists do not correspond in |
| length. |
Data corrupt or stale | |
| This occurs when a FETC? is attempted and either a reset has been |
| received or the power meter state has changed such that the current |
| measurement is invalidated (for example, a change of frequency setting |
| or triggering conditions). |