Power Meter Operation
Making Measurements using Frequency Dependent Offset Tables
| ■ All characters must be upper or lower case alphabetic |
|
|
| |||||||||
|
|
| characters, or numeric |
|
|
| |||||||
| ■ No other characters are allowed. |
|
|
| |||||||||
| ■ No spaces are allowed in the name. |
|
|
| |||||||||
| 2. To edit the parameter you currently have selected press |
|
| . | |||||||||
| Change | ||||||||||||
| The power meter displays the data in a pop up window. Modify |
|
|
| |||||||||
| this data (see below) as desired. |
|
|
| |||||||||
| ■ | Use |
| or | to modify the character on which the cursor | ||||||||
|
|
| is currently positioned. |
|
|
| |||||||
| ■ | Use |
| or | to move to other characters. |
|
|
| |||||
| 3. To confirm your choice press the appropriate softkey. |
|
|
| |||||||||
| 4. To adda new frequency dependent offset table entry, press |
|
|
| |||||||||
|
|
| . You are prompted for the frequency and offset. The entry | ||||||||||
|
| Insert | |||||||||||
| is sorted by frequency. |
|
|
| |||||||||
| 5. To remove a frequency dependent offset table entry, use the | , | |||||||||||
|
|
| , |
| and | hardkeys to select the entry, press |
|
|
| . If | |||
|
|
|
|
|
| ||||||||
|
|
|
|
| Delete |
| |||||||
| you delete the frequency the equivalent offset is also |
|
|
|
|
|
| ||||||
| removed and | ||||||||||||
| vice versa. |
|
|
|
|
|
|
|
|
| |||
|
|
|
|
|
|
|
|
|
|
| |||
Note | You can only edit the table name you cannot delete it. |
|
|
| |||||||||
|
|
|
|
|
|
|
|
|
| ||||
| 6. When you have completed editing the frequency dependent offset | ||||||||||||
| table, press |
| to return to the “Offset Tbls” screen. |
|
|
| |||||||
| Done |
|
|
| |||||||||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
Note | Ensure that the frequency points you use cover the frequency |
|
|
| |||||||||
| range of the signals you want to measure. If you measure a signal |
|
|
| |||||||||
| with a frequency outside the frequency range defined in the |
|
|
| |||||||||
| frequency dependent offset table, the power meter uses the |
|
|
| |||||||||
| highest or lowest frequency point in the frequency dependent |
|
|
| |||||||||
| offset table to calculate the offset. |
|
|
| |||||||||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
HP E4419B User’s Guide |